  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67263",
      "recalling_firm": "Synthes, Inc.",
      "address_1": "1302 Wrights Ln E",
      "address_2": "N/A",
      "postal_code": "19380-3417",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution-including the states of AR, AK, VA, IN, PA, OH, CT, TX, MI, WI, NC, CO, IL, and MA.",
      "recall_number": "Z-1274-2014",
      "product_description": "Drill Template and the Saw Guide for the Ulna Osteotomy System.    Intended for fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragments, including osteopenic bone in the ulna.",
      "product_quantity": "343",
      "reason_for_recall": "The Drill Template and the Saw Guide for the Ulna Osteotomy System may exhibit jamming, bending, or breaking of the screw connecting the Saw Guide to the Drill Template or jamming, bending, or breaking of the K-wire during fixation of the Drill Template.",
      "recall_initiation_date": "20140107",
      "center_classification_date": "20140325",
      "termination_date": "20150910",
      "report_date": "20140402",
      "code_info": "Part Nos. 03.111.900 through 03.111.905, with Lot Nos. 779372, 2638601, 7769386, 2638603, 7836420, 7772531, 2638604, 7836422, 7772538, 2638605, 7836423, 7915554, 7772543, 7772542, 7872775, 8400287, 7772554, 7772555, and 8229040."
    }
  ]
}