  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Blue Ash",
      "address_1": "4545 Creek Rd",
      "reason_for_recall": "The staplers may have an insufficient firing stroke to break the washer and completely form staples.",
      "address_2": "",
      "product_quantity": "434,614 total",
      "code_info": "All Lots within EXPIRATION Date Range:  February 2023   March 2024",
      "center_classification_date": "20190515",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Argentina, Aruba, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, El Salvador, Guatemala, Honduras, India, Indonesia, Israel, Jamaica, Japan, Jordan, Korea, Kuwait, Lebanon, Malaysia, Maldives, Mexico, Nepal. Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Qatar, Saudi Arabia, Singapore, Sri Lanka, Taiwan, Thailand, UAE, Uruguay, Vietnam, and Yemen.",
      "state": "OH",
      "product_description": "Endoscopic Curved Intraluminal Stapler, 25 mm diameter, Model ECS25A    The Ethicon Endo-Surgery Intraluminal Staplers (ILS) are anastomotic staplers available in four sizes to permit proper matching of instrument to diameter of the lumen. The instrument permits tissue attachment to the anvil shaft in a location remote from the main body of the instrument to improve access and visibility.",
      "report_date": "20190522",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Ethicon Endo-Surgery Inc",
      "recall_number": "Z-1272-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "82539",
      "more_code_info": "",
      "recall_initiation_date": "20190411",
      "postal_code": "45242-2803",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Completed"
    }
  ]
}