{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91752",
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "address_1": "700 Orthopaedic Dr",
      "address_2": "N/A",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Poland, Germany, India, Australia, & New Zealand",
      "recall_number": "Z-1267-2023",
      "product_description": "Attune Posterior (PS) Fixed Bearing (FB) Tibial Insert SZ5 7MM, Part Number 151640507",
      "product_quantity": "12 units",
      "reason_for_recall": "The subject product lots are being recalled because they received a higher than specified irradiation dose. This exceeds the validated range for exposure to gamma radiation of these devices and may result in changes to the implant material properties.",
      "recall_initiation_date": "20230215",
      "center_classification_date": "20230317",
      "report_date": "20230329",
      "code_info": "GTIN: 10603295049944; Lot: JN6613"
    }
  ]
}