  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Malvern",
      "address_1": "40 Liberty Blvd",
      "reason_for_recall": "When the archiving configuration is changed, data received/created after the upgrade may be flagged as \"Not to be archived\".  This is caused by the automatic function for cleaning up temporary data being disabled by the software upgrade. Due to the disabled cleanup function,  disc capacity for free space decreases faster than  usual. Unless the archiving configuration and the cleanup automatic function are reverted back to the original state (prior to the software upgrade), data that was incorrectly flagged \"Not to be archived\" must be manually prevented from being deleted by either clinical administrator or service engineer.",
      "address_2": "",
      "product_quantity": "165 units",
      "code_info": "All Syngo.via systems with VB20A. Model number 10496180.",
      "center_classification_date": "20180328",
      "distribution_pattern": "Medical device software which needs to be installed.",
      "state": "PA",
      "product_description": "Syngo.via software is intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a standalone device or together with a variety of cleared and unmodified syngo based software options.",
      "report_date": "20180404",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "recall_number": "Z-1261-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79513",
      "termination_date": "20181002",
      "more_code_info": "",
      "recall_initiation_date": "20180110",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}