{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91677",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide and Nationwide Distribution Foreign: Aregentina Brazil Chile Colombia Country India Japan New Zealand Nicaragua Saudi¿Arabia¿¿¿ South Africa South Korea Spain¿¿¿¿¿¿¿¿¿¿ Sri Lanka Taiwan",
      "recall_number": "Z-1258-2023",
      "product_description": "Dimension Tacrolimus (TAC) Flex reagent cartridge-In vitro diagnostic test for the quantitative measurement of tacrolimus in human whole blood on the Dimension¿ clinical chemistry system SMN: 10700795 (DF207)",
      "product_quantity": "6035 units",
      "reason_for_recall": "Siemens Healthineers has received customer complaints and confirmed imprecision for Quality Control (QC) and patient samples with Dimension Tacrolimus (TAC) lots GA2286, GA3047 and GA3171. The health consequences of erroneous patient results are major, since in a worst case scenario, a patient s tacrolimus dosing could be altered, resulting in potential organ failure and rejection (due to too low or no tacrolimus being administered) or toxicity (due to too much tacrolimus being administered). Siemens internal investigation of the Dimension TAC assay showed the worst case imprecision with patient samples at the low end of the Analytical Measurement Range (AMR). A patient sample at 2.0 ng/mL (2.6 nmol/L) recovered as 0.0 ng/mL (0.0 nmol/L) (100% negative bias), another patient sample at 2.1 ng/mL (2.7 nmol/L) recovered at 4.3 ng/mL (5.6 nmol/L)  (103% positive bias).",
      "recall_initiation_date": "20230130",
      "center_classification_date": "20230315",
      "report_date": "20230322",
      "code_info": "UDI: 00842768035425GA228622101310700795840, 00842768035425GA304723021610700795840, 00842768035425GA317123062010700795840. Lot Numbers: GA2286, GA3047, GA3171"
    }
  ]
}