  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Glendora",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84538",
      "recalling_firm": "Ormco/Sybronendo",
      "address_1": "1332 S Lone Hill Ave",
      "address_2": "",
      "postal_code": "91740-5339",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US:AK,AL,AR,AZ,CA,CO,CN,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV,WY    OUS: Argentina, Australia, Barbados, Bahamas, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Hong Kong, Indonesia, Japan, Cambodia, Korea, Cayman Islands, Sri Lanka, Myanmar, Mexico, Malaysia,  Netherlands, Nepal, Panama, Philippines, Paraguay, Russian Federation, Singapore, Thailand, Taiwan,  Uruguay, Viet Nam, South Africa",
      "recall_number": "Z-1258-2020",
      "product_description": "SnapLink, Ref Part No. 438-2191, L6L SnapLink, Slot .022, TQ--28, Ang +2, Rot +2, Qty 10, Made in Mexico, CE 0086, Rx Only",
      "product_quantity": "3,632,493 brackets",
      "reason_for_recall": "The action is being taken due to an increase in complaints relating to the lower buccal tube slide doors not staying close. In the event of functional failure, broken/damage buccal tubes it may cause the tooth to rotate, possibly delay treatment and inconvenience the patient and clinician. If not noticed it may extend treatment by approximately 1 -3 months.",
      "recall_initiation_date": "20200117",
      "center_classification_date": "20200218",
      "termination_date": "20240930",
      "report_date": "20200226",
      "code_info": "All Lots  UDI #s: 00889989002652 & 00889989002669",
      "more_code_info": ""
    }
  ]
}