  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90153",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Albania, Argentina, Austria, Bahamas, Bangladesh, Belgium, Brazil, Chile, China, Columbia, Croatia, Czech¿Republic, Denmark, Ecuador, Egypt, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Latvia, Malaysia, Mexico, Netherlands, Paraguay, Peru, Philippines, Portugal, Romania,  Russian¿Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., United Kingdom, Vietnam.",
      "recall_number": "Z-1252-2022",
      "product_description": "Atellica IM 1300 Analyzer - automated, immunoassay analyzer designed to perform in vitro diagnostic tests on clinical specimens.  Siemens Material Number (SMN):\t11066001",
      "product_quantity": "1,566",
      "reason_for_recall": "Lots incompatible with Test Definition (TDef) Version 1.4,  potentially lead to the generation of erroneous total T3 results",
      "recall_initiation_date": "20220413",
      "center_classification_date": "20220609",
      "report_date": "20220615",
      "code_info": "UDI: 00630414007960  All lots",
      "more_code_info": ""
    }
  ]
}