      "status": "Ongoing",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93874",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.",
      "recall_number": "Z-1249-2024",
      "product_description": "MRI system: Vantage Elan, Model: MRT-2020",
      "product_quantity": "1",
      "reason_for_recall": "For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in \"System Information\" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.",
      "recall_initiation_date": "20240129",
      "center_classification_date": "20240304",
      "report_date": "20240313",
      "code_info": "UDI-DI: 04987670101632. Serial Number: S1F2052753",
      "more_code_info": ""
    }
  ]
}