  "results": [
    {
      "status": "Ongoing",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90146",
      "recalling_firm": "Getinge Usa Sales Inc",
      "address_1": "1 Geoffrey Way",
      "address_2": "",
      "postal_code": "07470-2035",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of MD, MN, ND, NE, NY, PA, VA.",
      "recall_number": "Z-1248-2022",
      "product_description": "Getinge CM320-2 CHAMBER WU Series Washer Disinfector- For Use as a multi-chamber washer-disinfector for the washing, disinfecting and drying surgical instruments, etc.  Model: CM320 WU",
      "product_quantity": "4 units",
      "reason_for_recall": "Verification of installation not documented or may be incomplete, could result in harm to the user such as skin burns, smoke inhalation, electric shock, or a crush or pinch injury or slip and fall injury. May lead to a patient procedural delay and/or infection",
      "recall_initiation_date": "20220413",
      "center_classification_date": "20220608",
      "report_date": "20220615",
      "code_info": "UDI-DI: 07340153700185 Serial Numbers: W50034570 WAA070170 WAA070618 WAA081312",
      "more_code_info": ""
    }
  ]
}