  "results": [
    {
      "status": "Ongoing",
      "city": "Saint Charles",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98255",
      "recalling_firm": "Vortex Surgical Inc.",
      "address_1": "4 Research Park Dr Ste 124",
      "address_2": "N/A",
      "postal_code": "63304-5639",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.",
      "recall_number": "Z-1245-2026",
      "product_description": "Vortex Surgical    1. 23GA Laser Probe Curved  Cat No. VS0120.23    2. 25GA Laser Probe Curved  Cat No. VS0120.25    3. 25GA Laser Probe Flex-Tip  Cat No. VS0130.25    4. 25GA Laser Probe Illuminated Curved  Cat No. VS0125.25B    5. 25GA Laser Probe MaxReach  Cat No. VS0140.25    6. 25GA Laser Probe Straight  Cat No. VS0110.25",
      "product_quantity": "14,789 (8651 US; 6138 OUS)",
      "reason_for_recall": "There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.",
      "recall_initiation_date": "20251216",
      "center_classification_date": "20260204",
      "report_date": "20260211",
      "code_info": "1. 23GA Laser Probe Curved Pouch UDI 810123480036 Box UDI 810123480180 Lot 2411027  2. 25GA Laser Probe Curved Pouch UDI 810123480043 Box UDI 810123480197 Lot 2411026  3. 25GA Laser Probe Flex-Tip Pouch UDI 810123480111 Box UDI 810123480265 Lot 2411024 2411031  4. 25GA Laser Probe Illuminated Curved Pouch UDI 810123480098 Box UDI 810123480241 Lot 2411025  5. 25GA Laser Probe MaxReach Pouch UDI 810123480135 Box UDI 810123480289 Lot 2410029 2411023  6. 25GA Laser Probe Straight Pouch UDI 810123480012 Box UDI 810123480166 Lot 2411030"
    }
  ]
}