  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Alameda",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64876",
      "recalling_firm": "Abbott Diabetes Care, Inc.",
      "address_1": "1360 S Loop Rd",
      "address_2": "N/A",
      "postal_code": "94502-7000",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Press Release",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-1245-2013",
      "product_description": "FreeStyle InsuLinx Blood Glucose Monitoring System:  For In Vitro Diagnostic Use only. Use only with FreeStyle InsuLinx test strips.    Abbott Diabetes Care, Inc.  1360 South Loop Road  Alameda, CA 94502 USA    The FreeStyle InsuLinx Blood Glucose Test Strips are for use with the FreeStyle InsuLinx Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip.",
      "product_quantity": "335,273 meters; 209,336 remain at firm.",
      "reason_for_recall": "AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading.",
      "recall_initiation_date": "20130415",
      "center_classification_date": "20130509",
      "termination_date": "20150803",
      "report_date": "20130515",
      "code_info": "Part Numbers:  71142-70  71143-70  71145-70  71150-70:  All serial and lot numbers."
    }
  ]
}