  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "West Valley City",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96158",
      "recalling_firm": "Scientia Vascular, Inc.",
      "address_1": "2460 S 3270 W",
      "address_2": "N/A",
      "postal_code": "84119-1116",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U..S. Nationwide distribution in the states of AL, AZ, FL, GA, HI, IL, IN, KS, MN, NC, NJ, NY, PA, SD, TX, VA, WI and WV.",
      "recall_number": "Z-1234-2025",
      "product_description": "Socrates 38 Aspiration Catheter  127 cm length REF SC038-127-001    The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke",
      "product_quantity": "120 devices",
      "reason_for_recall": "Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.",
      "recall_initiation_date": "20250102",
      "center_classification_date": "20250226",
      "report_date": "20250305",
      "code_info": "Catalog Number: SC038-127-001 UDI-DI code: 00818075011025 Lot Numbers: 030471 030317 030345 030376 030427 030451"
    }
  ]
}