      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.",
      "address_2": "",
      "product_quantity": "3454",
      "code_info": "903213125  903213130  903213135",
      "center_classification_date": "20180327",
      "distribution_pattern": "USA (nationwide)",
      "state": "IN",
      "product_description": "ACE Trochanteric Nail System  TROCH NAIL ST 13X180 125DG STRL  TROCH NAIL ST 13X180 130DG STRL  TROCH NAIL ST 13X180 135DG STRL    The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal areas, long subtrochanteric fractures, ipsilateral femoral fractures, proximal or distal nonunions and malunions and revision procedures.",
      "report_date": "20180404",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-1231-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79408",
      "termination_date": "20200422",
      "more_code_info": "",
      "recall_initiation_date": "20171220",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}