  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91645",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide.  Global Distribution.",
      "recall_number": "Z-1229-2023",
      "product_description": "Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, and 0998-00-0800-55",
      "product_quantity": "8909 units",
      "reason_for_recall": "There have been reported failures of the high pressure helium regulator, which may cause a helium leak in the Cardiosave Hospital Cart.  In instance of helium regulator failure, a Pump Console s internal reservoir of helium will not be replenished when docked into an impacted Hospital Cart. This may result in an insufficient amount of helium within the internal reservoir.",
      "recall_initiation_date": "20230207",
      "center_classification_date": "20230306",
      "report_date": "20230315",
      "code_info": "All Serial Numbers.   Model Number: 0998-00-0800-31, UDI-DI:10607567109053; Model Number: 0998-00-0800-32, UDI-DI: 10607567111117; Model Number: 0998-00-0800-33, UDI-DI: 10607567109008; Model Number: 0998-00-0800-34, UDI-DI: 10607567111940; Model Number: 0998-00-0800-35, UDI-DI: 10607567109107; Model Number: 0998-00-0800-45, UDI-DI: 10607567108421; Model Number: 0998-00-0800-52, UDI-DI: 10607567108438; Model Number: 0998-00-0800-53, UDI-DI: 10607567108391; Model Number: 0998-00-0800-55, UDI-DI: 10607567108414; Model Number: 0998-00-0800-65, UDI-DI: 10607567113432",
      "more_code_info": ""
    }
  ]
}