      "status": "Terminated",
      "city": "Marcy L'Etoile",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76329",
      "recalling_firm": "BioMerieux SA",
      "address_1": "Chemin De L'Orme",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US including AZ, AR, CA, CT, DE, DC, FL, GA, ID, IL, IN, IA, KY, LA, MD, MA, MI, MS, MO, NE, NH, NJ, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WV, WI, and Internationally to Algeria, United Arab Emirates, Austria, Australia, Brazil, Cambodia, Canada, Chile, Chine, Costa Rica, Croatia, Czech Republic, Ecuador, Estonia, France, Greece, Germany, Great Britain, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, Nigeria, Norway,  Oman, Pakistan, Philippines, Poland, Portugal, Qatar, Romania, Russia, San Marino, Saudi Arabia, Senegal, Serbia, Singapore,  Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, Uruguay, and Vietnam",
      "recall_number": "Z-1226-2017",
      "product_description": "ETEST XM256 (Cefuroxime), Foam: 506958-506918; SPB: 412304-412305    ETEST is a quantitative technique for determining the antimicrobial susceptibility of Gram-negative and Gram-positive aerobic bacterial such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria",
      "product_quantity": "7,160 units",
      "reason_for_recall": "Potential performance issues.  False Susceptible result instead of Intermediate and False Intermediate result with certain methods.",
      "recall_initiation_date": "20170118",
      "center_classification_date": "20170221",
      "termination_date": "20200130",
      "report_date": "20170301",
      "code_info": "REF 412304 - 1004497830; REF 412305 - 1002884720, 1003028270, 1003075500, 1003383650, 1004018820, 1004497810, 1005161160; REF 506918 - 1002884980, 1003028290, 1003075850, 1003385930, 1003396150, 1004022170, 1004498100, 1005161470;  REF 506958 - 1002885010, 1003075540, 103396550, 1004019050, 1004501930, 1005161480"
    }
  ]
}