  "results": [
    {
      "status": "Ongoing",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93925",
      "recalling_firm": "Ortho-Clinical Diagnostics, Inc.",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY,  LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.",
      "recall_number": "Z-1225-2024",
      "product_description": "VITROS 4600 Chemistry System Catalog Nos. 6802445 (regular) and 6900440 (refurbished)",
      "product_quantity": "4 US; 18 OUS",
      "reason_for_recall": "A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.",
      "recall_initiation_date": "20240126",
      "center_classification_date": "20240301",
      "report_date": "20240313",
      "code_info": "VITROS System Software Version 3.8.1 Regular UDI-DI 10758750012343 Refurbished UD-DI 10758750033201  Serial Numbers  US: J46000240 J46000898 J46001030 J46001293  OUS: J46000150 J46001041 J46000190 J46001042 J46000499 J46001186 J46000564 J46001187 J46000801 J46001210 J46000821 J46001217 J46000910 J46001289 J46000933 J46001639 J46001029 J46001645",
      "more_code_info": ""
    }
  ]
}