{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91610",
      "recalling_firm": "St. Jude Medical",
      "address_1": "387 Technology Cir Nw Ste 500",
      "address_2": "",
      "postal_code": "30313-2424",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.",
      "recall_number": "Z-1219-2023",
      "product_description": "CardioMEMS HF System Patient Electronics System (PES), Model number CM1010, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients.  The system measures changes in PA pressure which physicians use to initiate or modify heart failure treatment.",
      "product_quantity": "200 units",
      "reason_for_recall": "Radiated Radiofrequency (RF) emissions of the  CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).",
      "recall_initiation_date": "20230207",
      "center_classification_date": "20230303",
      "report_date": "20230315",
      "code_info": "UDI/DI 05414734509176, Serial Numbers: M140300439, M140300440, M140300441, M140300445, M140300446, M140300447, M140300449, M140300450, M140300451, M140300454, M140300458, M140300459, M140300460, M140300463, M140500480, M140500486, M140800987, M140800988, M140800992, M140800995, M140801002, M140801003, M140801004, M140801007, M140801010, M140801011, M141001528, M141001529, M141001530, M141001531, M141001536, M141001539, M141001541, M141001544, M141001545, M141001546, M141101678, M141101681, M141101682, M141101683, M141101684, M141101685, M141101686, M141101693, M141101696, M141101697, M141101700, M141101701, M141101703, M141101705, M141101706, M150102038, M150102039, M150102049, M150102052, M150102054, M150102055, M150102059, M150102061, M150102062, M150102064, M150102066, M150102399, M150102402, M150102404, M150102409, M150102410, M150102414, M150102416, M150202537, M150202538, M150202541, M150202542, M150202544, M150202545, M150302658, M150302660, M150302662, M150302664, M150302666, M150302667, M150302669, M150302673, M150302680, M150302682, M150302683, M150302687, M150302689, M150302690, M150302692, M150302694, M150302695, M150302702, M150302703, M150302707, M150302711, M150302712, M150302713, M150302716, M150302721, M150302723, M150302724, M150302725, M150302729, M150302737, M150302739, M150302742, M150302747, M150302748, M150302751, M150302755, M150302757, M150302758, M150302759, M150302762, M150302769, M150302773, M150302777, M150302778, M150302779, M150302780, M150302782, M150302783, M150302786, M150403048, M150403050, M150403051, M150403057, M150403066, M150805139, M150805143, M150805151, M150805154, M150805155, M150805156, M150805158, M150805160, M150805161, M150805162, M150805166, M150905677, M150905680, M150905682, M150905683, M150905685, M150905688, M150905689, M150905691, M150905692, M150905697, M150905698, M150905699, M150905702, M150905703, M150905704, M150905706, M150905707, M150905708, M150905709, M150905711, M150905714, M151206628, M151206631, M151206633, M151206636, M151206637, M151206641, M151206644, M151206645, M151206647, M151206648, M151206651, M151206655, M151206659, M151206662, M151206665, M160206849, M160206853, M160206854, M160206857, M160206861, M160206867, M160206869, M160206872, M160206876, M160206879, M160206880, M160206882, M160206885, M160206888, M160206896, M160206899, M160206900, M160206902, M160206903, M160206906, M160807699, M160807700, M160807701, M160807772.",
      "more_code_info": ""
    }
  ]
}