  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bethlehem",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96212",
      "recalling_firm": "Tyber Medical",
      "address_1": "83 S Commerce Way Ste 310",
      "address_2": "N/A",
      "postal_code": "18017-8934",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide",
      "recall_number": "Z-1213-2025",
      "product_description": "A.L.P.S. mvX- ANATOMIC LAT FIB PLATE 8H LT: Anatomic Lateral Fibula Plate, 8-hole, Left  Intended to bridge or otherwise stabilize bone fragments to facilitate healing.  Model/Catalog Number: 770708081",
      "product_quantity": "214 units",
      "reason_for_recall": "Complaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices",
      "recall_initiation_date": "20250121",
      "center_classification_date": "20250221",
      "report_date": "20250305",
      "code_info": "UDI-DI 00196449015566; Lot Numbers 265784, 24043BU01, 261290"
    }
  ]
}