  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Schwarzenbruck",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70545",
      "recalling_firm": "Iba Dosimetry Gmbh",
      "address_1": "Bahnhofstr. 5",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the states of AZ and OK., and the countries of  Germany, Austria, France, UK, Poland, Finland, The Netherlands, Spain, Turkey, Russia, and South Africa, Algeria, North and South America, Mexico, Brazil, Venezuela, Colombia, Panama, Asia, China, Japan, Philippines, Korea, India, Singapore, Thailand, Taiwan, Hong Kong, Malaysia and Guam.",
      "recall_number": "Z-1212-2015",
      "product_description": "COMPASS SW Version 3.1, Catalog Number CS10-100, medical linear accelerator, radiological",
      "product_quantity": "124 units",
      "reason_for_recall": "Error in the software.   During internal tests of the current development version of the Compass SW it was found that dose reconstruction for DMLC plans when the jaws move or mlc leaves reverse during beam on will not be scaled correctly for all control points.",
      "recall_initiation_date": "20150202",
      "center_classification_date": "20150226",
      "termination_date": "20150818",
      "report_date": "20150304",
      "code_info": "Model :  SW Version 3.1   Lot or Serial Number :  n.a",
      "more_code_info": ""
    }
  ]
}