  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Alameda",
      "address_1": "1360 S Loop Rd",
      "reason_for_recall": "Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with   Free Style test strips lots within expiry.",
      "address_2": "",
      "product_quantity": "not provided. Device has not been shipped in the last 5 years.",
      "code_info": "Part number CAT11001-0; all serial numbers",
      "center_classification_date": "20140318",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.",
      "state": "CA",
      "product_description": "FreeStyle Blood Glucose Monitors System.  For in vitro diagnostic use only. Store at room  temperature below 86 degrees F (30 degrees C). Use test strips  only within the system operating temperature  range as outlined in your Owner's Booklet.    Meter made in China;  Distributed by Abbott Diabetes Care, Alameda, CA    The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete can product inaccurate results.    The FreeStyle Glucose meter is incorporated into the OmniPod insulin Management system (manufactured by Insulet Corporation, Bedford, MA and is intended for subcutaneous delivery of insulin.    Intended to monitor blood glucose from samples taken from the body. IVD use only,",
      "report_date": "20140326",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Abbott Diabetes Care, Inc.",
      "recall_number": "Z-1211-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67472",
      "termination_date": "20150811",
      "more_code_info": "",
      "recall_initiation_date": "20140219",
      "postal_code": "94502-7000",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}