  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "325 Corporate Drive",
      "reason_for_recall": "Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect.  The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.",
      "address_2": "",
      "product_quantity": "107 units",
      "code_info": "Catalog number 703615",
      "center_classification_date": "20130501",
      "distribution_pattern": "Nationwide Distribution including the states of AL, AZ, IL, KY, LA,  NJ,  NC, FL, CA,  NM,  MO, TN, VA, CO, MA, TX,  IN and MT.",
      "state": "NJ",
      "product_description": "Stryker Howmedica Osteonics Corp.  AxSOS  System Instrument.    The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, \" A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve.\"",
      "report_date": "20130508",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Howmedica Osteonics Corp.",
      "recall_number": "Z-1210-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64942",
      "termination_date": "20151117",
      "more_code_info": "",
      "recall_initiation_date": "20130328",
      "postal_code": "07430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}