  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Montreal",
      "state": "N/A",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96270",
      "recalling_firm": "INTELERAD MEDICAL SYSTEMS INCORPORATED",
      "address_1": "800, Boul. de Maisonneuve Est",
      "address_2": "12th floor",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia",
      "recall_number": "Z-1208-2025",
      "product_description": "IntelePACS (Image Fusion Module) - InteleViewer",
      "product_quantity": "88",
      "reason_for_recall": "A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.",
      "recall_initiation_date": "20250205",
      "center_classification_date": "20250221",
      "report_date": "20250305",
      "code_info": "UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later"
    }
  ]
}