  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cuyahoga Falls",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91601",
      "recalling_firm": "Coltene Whaledent Inc",
      "address_1": "235 Ascot Pkwy",
      "address_2": "N/A",
      "postal_code": "44223-3701",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide Foreign: Canada Germany Indonesia Ireland Lebanon Saudia Arabia Turkey",
      "recall_number": "Z-1208-2023",
      "product_description": "SciCan STATIS SL S10001 Dental Handpieces-Intended for the removal of carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations, restorations and for polishing teeth.  Model Number: S10001",
      "product_quantity": "3896 units",
      "reason_for_recall": "Screw connection between the turbine head and the push button may loosen and fall off during the treatment, and increase the risk of swallowing or inhaling part of the turbines head or the push button",
      "recall_initiation_date": "20230120",
      "center_classification_date": "20230302",
      "report_date": "20230308",
      "code_info": "UDI-DI: +ESANS100010  All Serial Numbers",
      "more_code_info": ""
    }
  ]
}