  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Reading",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84726",
      "recalling_firm": "Arrow International Inc",
      "address_1": "2400 Bernville Rd",
      "address_2": "",
      "postal_code": "19605-9607",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The products were distributed to the following US states:  AK, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY.",
      "recall_number": "Z-1199-2020",
      "product_description": "EPIDURAL CATHETERIZATION KIT, Product Code FR-05501-12",
      "product_quantity": "343,105 total",
      "reason_for_recall": "There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.",
      "recall_initiation_date": "20200115",
      "center_classification_date": "20200213",
      "termination_date": "20230317",
      "report_date": "20200219",
      "code_info": "Lots: 71F19A0623  71F19C1042  71F19E1144  71F19G2853  71F19K0248  71F19A2996  71F19C2596  71F19E2950  71F19H2549  71F19K2117  71F19B1149  71F19D2545  71F19G1138  71F19J1815  71F19L0436",
      "more_code_info": ""
    }
  ]
}