{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Minnetonka",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96155",
      "recalling_firm": "Augustine Temperature Management, LLC",
      "address_1": "15305 Minnetonka Blvd",
      "address_2": "N/A",
      "postal_code": "55345-1512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution in the states of TX, AL, NV; AU, NZ.",
      "recall_number": "Z-1197-2025",
      "product_description": "Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32\"), REF US501, to prevent or treat hypothermia",
      "product_quantity": "4 units",
      "reason_for_recall": "There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It  turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.",
      "recall_initiation_date": "20241216",
      "center_classification_date": "20250220",
      "report_date": "20250226",
      "code_info": "UDI-DI UDI-DI 00855913001817, Serial Numbers:  1304292, 1304293, 1304296, 1304297"
    }
  ]
}