  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tustin",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "97970",
      "recalling_firm": "Canon Medical System, USA, INC.",
      "address_1": "2441 Michelle Dr",
      "address_2": "N/A",
      "postal_code": "92780-7047",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Dominican Republic.",
      "recall_number": "Z-1177-2026",
      "product_description": "Alphenix INFX-8000F, interventional fluoroscopic x-ray system",
      "product_quantity": "98 units",
      "reason_for_recall": "It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional   systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed",
      "recall_initiation_date": "20251219",
      "center_classification_date": "20260127",
      "report_date": "20260204",
      "code_info": "serial numbers:  W1C0582204, B2A1582208, W2A1052118, A3582035, A2A2362689, B2B18Y2813, W1D07Y2447, B4582113, W2A1042111, W1C0622270, A3592048, W4A11X2138, W2A09Y2060, B2A16Z2476, W2A09Y2066, W4A1122041, B2A1482032, B2A1482027, A2A2252435, W1D0722369, B2B15Y2247, W1B0542168, B2A1492049, W1D0722371, A2A2032031, A1C1972103, W4C1352375, B2B1862737, W1D0842499, W4A1212179, W4A11Z2171, W4A11Z2173, W1C0652298, B2A1692419, B2A1552170, A2A21Y2350, W4A1242226, B2B1952848, A1C19Z2180, B2A1662371, W1D0892555, W1D0822476, W1D06Z2356, B2A1682394, B2A17Y2619, B2A17Y2624, W1D0762406, A2A2172283, W4A11Y2155, A2A2272460, W1D0832487, B2B1822667, A1C1992146, W4A1112031, W4C1382409, A2A2292508, W4A13X2015, W4A1192130, B2A1712510, B2A1762562, W4A1142066, W1C0622274, A4552098, W4A12X2298, W1D0812460, A4542088, W1D0712363, A2A21X2348, W1C0552179, W4A1142065, B2A17Z2626, W4D13Y2444, W2A1042110, A2A20Z2159, W4A11Y2147, W1C0642289, B2B1852720, B2A1592219, W4B13Y2016, B2B1922843, B2B1822668, W4B12Y2312, A2A2052056, A2A2062062, A2A2072075, A2A2042041, A1C19Z2181, W2A1092163, A2A2012005, A2A20Y2147, A1C19Y2161, B2A1612289, W4A1112030, W2A0982042, B2B18X2802, B2A1712503, B2A1712499, B2A14Z2084."
    }
  ]
}