{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Wayne",
      "address_1": "45 Barbour Pond Drive",
      "reason_for_recall": "When using the SERVO-i on battery power, in a very small number of cases, some battery modules distributed after January 31, 2010 have a shorter battery run time than expected.",
      "address_2": "",
      "product_quantity": "90,000 units (batteries)",
      "code_info": "SERVO-i Battery Module part number is 6487180.  Notice pertains to Battery Modules manufactured between February 2010 and October 2012 with date codes between 1005 and 1243.",
      "center_classification_date": "20130426",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) and countries of: ALGERIA, ANDORRA, ANTILLES (Netherlands), ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, AZERBADJAN, BANGLADESH, BELARUS, BELGIUM, BOLIVIA, BOTSWANA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA,  COLOMBIA, CONGO, COSTA RICA, CROATIA, CUBA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR,   ESTONIA, YUGOSLAVIA (Former), FIJI, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONDURAS, HONG KONG,  HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LIBYA, LITHUANIA, LUXEMBURG, MACAU, MACEDONIA, MALAYSIA, MALDIVES, MARTINIQUE (French), MEXICO, MOLDAVIA, MOROCCO, NEPAL, NETHERLANDS, NEW CALEDONIA (French), NEW ZEALAND, NICARAGUA,   NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PANAMA, PAPUA NEW GUINEA, PARAGUAY, PERU, PHILIPPINES, POLAND, POLYNESIA (French),   PORTUGAL, QATAR, REUNION (French), ROMANIA, RUSSIA, SAUDI ARABIA,   SERBIA, SEYCHELLES, SINGAPORE, SLOVAK REPUBLIC, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SUDAN, SURINAM, SWEDEN, SWITZERLAND, SYRIA, TADJIKISTAN, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VATICAN city state, VENEZUELA, VIETNAM and YEMEN.",
      "state": "NJ",
      "product_description": "Maquet Getinge Group battery modules for use with Maquet SERVO-i ventilator systems (Part #6487180).    Battery for use with the SERVO-i Ventilator System. The SERVO-i Ventilator System is intended for treating and monitoring patients ranging from neonates to adults with respiratory failure or insufficiency. The ventilator is equipped with at least 2, NiMH battery modules (Part No. 6487180) which automatically supply 12V DC Power in case of an AC power failure, ensuring that ventilator settings and stored data remain intact in the event of an AC power failure. The battery modules give the user a possibility to use the SERVO-i during intra-hospital transportation.",
      "report_date": "20130508",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "recall_number": "Z-1177-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64645",
      "termination_date": "20160811",
      "more_code_info": "",
      "recall_initiation_date": "20130307",
      "postal_code": "07470",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}