      "country": "United States",
      "city": "Costa Mesa",
      "address_1": "1620 Sunflower Ave",
      "reason_for_recall": "Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70  and HT70 Plus ventilator Power Pac batteries due to customer reports of Newport HT70 and HT70 Plus ventilators alarming and going to internal backup battery sooner than expected while the ventilator is being operated on Power Pac battery.",
      "address_2": "",
      "product_quantity": "2528 ventilators and 1864 power pac batteries",
      "code_info": "Model number: HT70 and HT70 Plus    Power Pac battery serial number range: 2096134110001 to  2292048120050, Rev. C through Rev. F",
      "center_classification_date": "20130426",
      "distribution_pattern": "Worldwide Distribution - USA including US (AL, AR, AZ, CA, CO, CT, DE, FL, GA, lA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, TN, TX, UT, VA, WA, and WI. Internationally to Argentina, Australia, Bangladesh, Belgium, BoliVia, Brazil,  Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, Egypt, France, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Ireland, Italy, Japan, Kazakhstan, Kosova, Lebanon, Lithuania, Malaysia, Mexico, Nambia, Nepal, Norway, Oman, Paraguay, Peru, Philippines, Poland, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Sri Lanka, Switzerland,  Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Viet Nam.",
      "state": "CA",
      "product_description": "Newport Medical Instruments HT70  and HT70 Plus ventilator Power Pac batteries  Model number: HT70 and HT70 Plus.     The HT70 and HT70 Plus ventilator systems are intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces.",
      "report_date": "20130508",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Newport Medical Instruments Inc",
      "recall_number": "Z-1171-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64851",
      "termination_date": "20150310",
      "more_code_info": "",
      "recall_initiation_date": "20130405",
      "postal_code": "92626-1513",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}