{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78999",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide, Algeria, Australia, Austria,  Belarus, Belgium, Bosnia and Herzegovina,  Brazil, Canada, China,  Croatia, Cyprus, Czech Republic,  Denmark, Finland, France,  Germany, Gibraltar, Greece,  Hong Kong, Hungary, India,  Ireland, Israel, Italy,  Korea, Kuwait, Latvia,  Lithuania, Macao, Mayotte,  Netherlands, New Zealand, Norway,  Panama, Poland, Portugal,  Russia, Saudi Arabia, Slovakia,  Slovenia, South Africa, Spain,  Sweden, Switzerland, Taiwan,  Turkey, and United Kingdom.",
      "recall_number": "Z-1151-2018",
      "product_description": "AutoMate 2550  \"High Speed\"  Catalog Number ODL25255; AutoMate 1250  \"Standard\" Catalog Number ODL25125.  Chemistry analyzer sample processing and sorting system.",
      "product_quantity": "73 units in US",
      "reason_for_recall": "A misaligned pipette tip sensor poses the remote risk of cross-contamination when failing to cause a no-tip error, resulting in pipetting patient samples without pipette tips.",
      "recall_initiation_date": "20180118",
      "center_classification_date": "20180321",
      "termination_date": "20230927",
      "report_date": "20180328",
      "code_info": "All units in the AutoMate 2500 Family are affected; Serial Numbers - 1107,  4048,  4055,  4025,  4052,  4079,  4091,  4076,  4072,  4073,  4074,  4111,  4097,  1054,  1055,  1048,  1049,  1147,  ODL25125-1342,  ODL25125-1354,  1251,  ODL25255-4530,  ODL25255-4531,  4126,  4127,  4169,  4171,  4341,  4324,  ODL25255-4536,  ODL25255-4535,  ODL25255-4397,  4141,  4143,  4144,  4146,  ODL25255-4606,  4148,  ODL25255-4577,  ODL25255-4605,  4302,  4304,  4352,  ODL25255-4352,  4330,  ODL25255-4419,  ODL25255-4534,  ODL25255-4416,  ODL25255-4401,  4168,  4107,  4255,  4257,  4252,  4256,  4254,  4253,  4236,  ODL25255-4437,  ODL25255-4443,  ODL25255-4444,  ODL25255-4438,  4294,  4293,  ODL25255-4556,  ODL25255-4382,  ODL25255-4378,  ODL25255-4377,  ODL25255-4379,  ODL25255-4380,  ODL25255-4381,  ODL25255-4420,  ODL25255-4421"
    }
  ]
}