  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76275",
      "recalling_firm": "Vertebral Technologies, Inc.",
      "address_1": "13845 Industrial Park Blvd",
      "address_2": "N/A",
      "postal_code": "55441-3746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US: IL, WI, OH, NY, MT, FL, MD, IN, TN, TX, MN.    OUS: Italy, Taiwan Spain, Belgium, Canada, Netherlands, Sweden, Switzerland, Portugal, Brazil. Germany, Ireland,",
      "recall_number": "Z-1151-2017",
      "product_description": "InterFuse Tail Traction Tool, Part Number 9116.  The InterFuse Tail Traction Tool is an optional Class I non-sterile reusable instrument that can be used to assist with module locking of the InterFuse S or InterFuse T implant system.    The system has a modular design to allow the device to be assembled within the disc space to minimize the surgical window, while providing a footprint that approximates the excised nucleus.",
      "product_quantity": "63 pieces",
      "reason_for_recall": "Fit issues have been identified with the Tail Traction Tool (PN 9116) when used with the Interfuse T tailed implant.  The InterFuse T Tail may not fit into the Thru Hole in the instrument.",
      "recall_initiation_date": "20170113",
      "center_classification_date": "20170208",
      "termination_date": "20171005",
      "report_date": "20170215",
      "code_info": "Lot:  110512-02, 110718-01, 110831-02, 120214-01, 120403-018, 120420-07, 120711-02, 120911-06"
    }
  ]
}