  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Draper",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67641",
      "recalling_firm": "Megadyne Medical Products, Inc.",
      "address_1": "11506 South State Street",
      "address_2": "N/A",
      "postal_code": "84020-9453",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "USA including  FL, MN, LA, NJ, NM, PA, TX and Internationally to Canada.",
      "recall_number": "Z-1151-2014",
      "product_description": "Hand Control 12FR Suction Coagulator with 10 ft cord and holster, Sterile, Disposable, Model Number 0043-25, packaged Tyvek¿ polyester peel pouch, 25 pouches per box.    This device is intended to conduct monopolar electrosurgical energy from an electrosurgical generator (ESU) to target tissue during ENT and general surgical procedures. It is also a suction apparatus used to remove surgical smoke and fluids from the site. The device is intended for single use; it is not intended to be cleaned or reused.",
      "product_quantity": "494",
      "reason_for_recall": "Megadyne Medical Products, Inc. is voluntarily recalling Hand Control 12Fr Suction Coagulator with 10ft cord and 132405 Holster, Sterile, Disposable, Lot 132405 due to the potential for a small number of these suction coagulators to self-activate immediately on connection to the electrosurgical generator  and to remain activated until it is disconnected.",
      "recall_initiation_date": "20140227",
      "center_classification_date": "20140307",
      "termination_date": "20140709",
      "report_date": "20140319",
      "code_info": "Catalog Number: 0043-25  Lot Number: 132405  Exp. Date: 07-2018"
    }
  ]
}