{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91602",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, TURKMENISTAN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, AND YEMEN.",
      "recall_number": "Z-1146-2023",
      "product_description": "Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85",
      "product_quantity": "4062 Total",
      "reason_for_recall": "An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable  and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.",
      "recall_initiation_date": "20230207",
      "center_classification_date": "20230309",
      "report_date": "20230315",
      "code_info": "All serial numbers.  Model Number 0998-00-0800-75, UDI-DI 10607567112312; Model Number 0998-00-0800-83, UDI-DI  10607567108407; Model Number 0998-00-0800-85, UDI-DI  10607567113449",
      "more_code_info": ""
    }
  ]
}