  "results": [
    {
      "status": "Terminated",
      "city": "Eden Prairie",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64840",
      "recalling_firm": "Acist Medical Systems",
      "address_1": "7905 Fuller Rd",
      "address_2": "N/A",
      "postal_code": "55344-2137",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including the states of AL, AK, AZ, AR, CA, CO, CT, DC, GA, HI, ID, IL, IN, IA, KS, KY, MD,MI, MN, MS, MO, NV, NH, NJ, NM, NY,NC, OR, OK, OH, PA, TN, TX, VA and WA.",
      "recall_number": "Z-1145-2013",
      "product_description": "ACIST Medical Systems Inc.    The ACIST Angiographic Injection System is intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.",
      "product_quantity": "6640 units",
      "reason_for_recall": "ACIST Medical Systems is issuing a voluntary recall notice for three specific lots of BT2000 Automated Manifold Kits. For these specific lots, ACIST has received field reports related to ingress of air into the manifold body. This has been detected by the Air Column Detect Sensor halting the injection, and normally replacement of the BT2000 Automated Manifold Kit is necessary in order to proceed w",
      "recall_initiation_date": "20130315",
      "center_classification_date": "20130418",
      "termination_date": "20130807",
      "report_date": "20130424",
      "code_info": "23884007, 23884008, 23884009"
    }
  ]
}