  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87169",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1144-2021",
      "product_description": "ARTIS Icono biplane Model:11327600  ARTIS Icono floor system Model: 11327700   with software version VE20B - Product Usage: Intended for single and biplane diagnostic imaging and interventional procedures.",
      "product_quantity": "41 U.S",
      "reason_for_recall": "Misleading error messages and a gap in the Operator Manual which affects ARTIS Icono biplane or ARTIS Icono floor systems with software version VE20B. Potential  issues include System error management, Erroneous error messages, Zoom/Pan Function, Grid Indication, and Coolant Level. May cause  procedures  to be terminated and performed on an alternative x-ray system",
      "recall_initiation_date": "20201218",
      "center_classification_date": "20210223",
      "termination_date": "20211117",
      "report_date": "20210303",
      "code_info": "Software version VE20B.  Serial Numbers:  Serial  180054  180062  180019  180065  180056  180057  180041  180084  180034  180072  180060  170026  180070  180048  180083  180080  180067  180040  180029  170023  180068  170025  180018  180047  180046  180085  180035  180081  170051  170027  180043  170053  180049  180077  180053  180055  180015  180014  180066  180082  180037"
    }
  ]
}