  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80640",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed only to China and UK.",
      "recall_number": "Z-1142-2019",
      "product_description": "Brilliance 16 Air, 728246    Product Usage:  Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories",
      "product_quantity": "23",
      "reason_for_recall": "Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.",
      "recall_initiation_date": "20180703",
      "center_classification_date": "20190413",
      "termination_date": "20200417",
      "report_date": "20190424",
      "code_info": "6143  5701  50069  6232  6563  50007  5928  3466  6301  6328  50048  6145  5560  6324  5908  6520  5924  5496  6507  5944  6351  6149  6562"
    }
  ]
}