      "status": "Terminated",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82459",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of  CA, FL, MI, OR, PA, TX and WA.",
      "recall_number": "Z-1141-2019",
      "product_description": "Foot Switches used with the following systems:   722001 \tAllura Xper FD 10 C  722002\tAllura Xper FD10  F  722003\tAllura Xper FD10  722005\tAllura Xper FD10/10  722006\tAllura Xper FD20  722008\tAllura Xper FD20 Biplane  722010\tAllura Xper FD10  722011\tAllura Xper FD10/10  722012\tAllura Xper FD20  722013\tAllura XPER FD20 BIPLANE  722014\tAllura Xper FD10 OR Table  722015\tAllura Xper FD20 OR Table  722019\tAllura Xper FD10/10 OR Table  722020\tAllura Xper FD20 Biplane OR Table  722022\tAllura Xper FD10 OR Table  722023\tAllura Xper FD20 OR Table  722024\tAllura Xper FD10/10 OR Table  722025\tAllura Xper FD20 Biplane OR Table  722026\tAllura Xper FD10  722027\tAllura Xper FD10/10  722028\tAllura Xper FD20  722029\tAllura Xper FD20/10  722031\tAllura CV20  722033\tAllura Xper FD10 OR Table  722034\tAllura Xper FD10/10 OR Table  722035\tAllura Xper FD20 OR Table  722036\tAllura Xper FD20/10 OR Table  722038\tAllura Xper FD20/20  722039\tAllura Xper FD20/20 OR Table  722058\tAllura Xper FD20/15  722059\tAllura Xper FD20/15 OR Table  722123\tField extensions Xper cardio  722124\tField extensions Xper vascular  722126\tSmartPath to upgrade Allura to AlluraClarity cardio  722127\tSmartPath to upgrade Allura to AlluraClarity vascular  722133\tField ext. Xper cardio systems R7.6  722134\tField ext. Xper vascular systems  722400\tCardio Vascular-Allura Centron  722030\tINTEGRIS CV  722043\tINTEGRIS Allura 15-12 (mono)    Product The Allura Xper FD Series are intended for use on human patients to perform:   Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.",
      "product_quantity": "7209 total",
      "reason_for_recall": "Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.",
      "recall_initiation_date": "20190219",
      "center_classification_date": "20190413",
      "termination_date": "20201014",
      "report_date": "20190424",
      "code_info": "Foot Switches:  452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m;  452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2;  459800415531 Wireless Footswitch 3P;  459800415561 Wireless footswitch set 3P."
    }
  ]
}