      "status": "Terminated",
      "city": "Best",
      "state": "N/A",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82459",
      "recalling_firm": "Philips Medical Systems Nederlands",
      "address_1": "Veenpluis 4-6",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of  CA, FL, MI, OR, PA, TX and WA.",
      "recall_number": "Z-1140-2019",
      "product_description": "Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038    Product The Allura Xper FD Series are intended for use on human patients to perform:   Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.",
      "product_quantity": "7209 total",
      "reason_for_recall": "Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.",
      "recall_initiation_date": "20190219",
      "center_classification_date": "20190413",
      "termination_date": "20201014",
      "report_date": "20190424",
      "code_info": "Foot Switches:  452270000141 Footswitch CV 3p 4m; 452270000151 Footswitch MD 3p 6m;  452270000142 Footswitch CV 3p 4m; 452270000381 Footswitch CV 3p 8m; 452270000382 Footswitch CV 3p 8m; 459800076001 Footswitch CV 4p+2 4m; 459800076021 Footswitch CV 4p+2 8m; 459800415571 Wireless Footswitch 4p+2; 459800415581 Wireless footswitch set 4p+2;  459800415531 Wireless Footswitch 3P;  459800415561 Wireless footswitch set 3P."
    }
  ]
}