      "status": "Terminated",
      "city": "Marietta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "90022",
      "recalling_firm": "ProgenaCare",
      "address_1": "2275 Northwest Pkwy Se Ste 175",
      "address_2": "",
      "postal_code": "30067-9321",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AL, GA, FL, TN, and TX.  There was no foreign/military/government distribution.",
      "recall_number": "Z-1137-2022",
      "product_description": "progenamatrix Human Keratin Matrix, sterile, Rx, with the following model numbers: (1) PM2012, 12 x 2 cm; (2) PM3002, 2 x 2 cm; (3) PM3004, 4 x 4 cm; PM3006, 6 x 6 cm; and PM3010, 10 x 10 cm.",
      "product_quantity": "547 packages",
      "reason_for_recall": "Inner pouch seal may be inadequate resulting in lack of sterility.",
      "recall_initiation_date": "20220418",
      "center_classification_date": "20220526",
      "termination_date": "20230414",
      "report_date": "20220601",
      "code_info": "(1) Model PM2012, lot #02SEP21-001, exp. 4/9/2024; lot #29SEP21-001, exp. 4/9/2024.  (2) Model PM3002, lot #28OCT21-001, exp. 3/7/2024.  (3) Model PM3004, lot #02DEC21-001, exp. 10/1/2023; lot #02SEP21-002, exp. 10/1/2023; lot #14OCT21-001, exp. 10/1/2023; and 23NOV21-001, exp. 10/1/2023.  (4) Model PM3006, lot #02SEP21-003, exp. 10/1/2023; lot #03DEC21-001, exp. 10/1/2023; lot #14OCT21-002, exp. 10/1/2023; and lot #23NOV21-002, exp. 10/1/2023.  (5) Model PM3010, lot #13JUL21-001, exp. 10/1/2023; and lot #15JUL21-002, exp. 3/6/2024.",
      "more_code_info": ""
    }
  ]
}