      "status": "Terminated",
      "city": "Glens Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75226",
      "recalling_firm": "Angiodynamics Inc. (Navilyst Medical Inc.)",
      "address_1": "10 Glens Falls Tech Park",
      "address_2": "N/A",
      "postal_code": "12801-3864",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "AZ, CA, FL, GA, IL, IN, KS, MD, MO, MT, NC, NJ, NV, OH, OR, PA, TN, TX, WI",
      "recall_number": "Z-1137-2017",
      "product_description": "Celerity ECG Cable Accessory Pack, STERILE, Rx ONLY - distributed under the following labels: (a) angiodynamics UPN/Product No. H787471431, REF/Catalog No. 47-143 -- Legal Manufacturer: AngioDynamics, Inc. 10 Glens Falls Technical Park, Glens Falls, NY 12801; and (b) medCOMP UPN/Product No. MRLCA000, REF/Catalog No. MRLCA000 -- medCOMP, 1499 Delp Drive, Harleysville, PA 19438 --- Description: The Celerity ECG Cable Accessory Pack is comprised of two sterile, single use components: a cable cover and an ECG clip cable (alligator clip).  The cable cover is used to sleeve the remote control cable component of the Celerity System so that the user can remotely control the monitor within the sterile field.  The stereo jack end of the remote control cable pierces the sterile cable cover and interfaces with the stereo jack connector end of the ECG clip cable.  The alligator clip side of the ECG clip cable is clamped to the stylet or guidewire to provide the intravascular cavity ECG reading.",
      "product_quantity": "N/A",
      "reason_for_recall": "The Celerity ECG Cable Accessory Pack has the potential to contain an ECG Clip Cable that was not properly manufactured.  The affected product is not soldered per specification, and may result in the device not properly conducting the ECG signal.",
      "recall_initiation_date": "20160907",
      "center_classification_date": "20170203",
      "termination_date": "20170406",
      "report_date": "20170215",
      "code_info": "AngioDynamics UPN H787471431 Batch/Lot Numbers: 5015180, MBVR240D, MBVT150R, MBVT210D, MBVT220D, MBVT240D, MBVT250D, MBVT680D, MBVT700D, MBVX090D;  medCOMP UPN MRLCA000, Batch/Lot Number MBVT240."
    }
  ]
}