  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Jena",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87055",
      "recalling_firm": "Carl Zeiss Meditec AG",
      "address_1": "Carl-Zeiss-Promenade 10",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TX, UT, VA, VT, WA, WI, and WV.    O.U.S.: Austria, Australia, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, France, United Kingdom, Guatemala, Hong Kong, Indonesia, Ireland, India, Italy, Japan, Kuwait, Malaysia, Netherlands, New Zealand, Philippines, Poland, Portugal, Russia, Sweden, Singapore, Slovakia, Thailand, Taiwan, and South Africa.",
      "recall_number": "Z-1133-2021",
      "product_description": "IOLMaster 700",
      "product_quantity": "613 devices",
      "reason_for_recall": "When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen",
      "recall_initiation_date": "20201207",
      "center_classification_date": "20210218",
      "termination_date": "20220624",
      "report_date": "20210224",
      "code_info": "Model: IOLMaster 700  Catalog Number device: 000000-1932-169"
    }
  ]
}