  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82469",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to one account in Florida, and one international account in Germany.",
      "recall_number": "Z-1130-2019",
      "product_description": "Azurion FlexArm, 722079",
      "product_quantity": "2",
      "reason_for_recall": "During manufacturing of an Azurion R2.0 system, it was found out that the cable set was not according to specification. This resulted in damage of the 24V power cable of the detector.",
      "recall_initiation_date": "20190312",
      "center_classification_date": "20190411",
      "termination_date": "20201007",
      "report_date": "20190417",
      "code_info": "Serial numbers 637, 638"
    }
  ]
}