  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Chesterfield",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82497",
      "recalling_firm": "ulrich medical USA Inc",
      "address_1": "18221 Edison Ave",
      "address_2": "N/A",
      "postal_code": "63005-3703",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "AZ, IL, LA, MI, MO, NV, OH, PA, TN, and TX.",
      "recall_number": "Z-1126-2019",
      "product_description": "ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.",
      "product_quantity": "25",
      "reason_for_recall": "Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.",
      "recall_initiation_date": "20190308",
      "center_classification_date": "20190411",
      "termination_date": "20200710",
      "report_date": "20190417",
      "code_info": "Model #s CS 7130  Serial #s  538436-02_70 (69 total) UDI Description 4052536123704, Model #s CS 7134 Serial #s  538438-02_70 (69 total) UDI Description 4052536123728, Model #s CS 7146 Serial #s  04-14-022-18_23 (6 total) UDI Description4052536078301, Model #s CS 7148 Serial #s  11-12-035-01_09, 03-12-006-24, and 03-12-006-25 (11 total)  UDI Description 4052536078318,  and Model #s CS 8049 Serial #s  J04192-31_40, J04192-056, J04192-064, H12015-031_6, and J01276-031_3, (21 total)  UDI Description 4052536022076."
    }
  ]
}