  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98128",
      "recalling_firm": "ICU Medical, Inc.",
      "address_1": "600 N Field Dr",
      "address_2": "",
      "postal_code": "60045-4835",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Philippines.",
      "recall_number": "Z-1118-2026",
      "product_description": "ICU Medical Plum Duo Infusion Pump, List Numbers:  400020401 (US), 400021001 (Philippines)",
      "product_quantity": "",
      "reason_for_recall": "Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.",
      "recall_initiation_date": "20251215",
      "center_classification_date": "20260116",
      "report_date": "20260128",
      "code_info": "400020401 (US), Software Versions:  1.1.1, 1.1.2, 1.1.3, 1.1.4 400021001 (Philippines), Software Versions:  1.1.0",
      "more_code_info": ""
    }
  ]
}