      "country": "United States",
      "city": "Ann Arbor",
      "address_1": "6200 Jackson Rd",
      "reason_for_recall": "These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed.  The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.",
      "address_2": "",
      "product_quantity": "16",
      "code_info": "Serial Numbers: 882298  882299  882300  882752  882753  882794  882795  882796  882797  882798  882799  885547  885548  885549  885550  885553",
      "center_classification_date": "20200207",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of CT, FL, LA, MA, MI, NY, and TX. Countries of Hong Kong, India, South Korea, and Taiwan.",
      "state": "MI",
      "product_description": "Terumo Advanced Perfusion Assembly Pump Guts, 6 Inch, Catalog Number 801798 - is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical proffessional who is experienced in the operation of this or similar equipment. The roller pump is used to move fluids through the cardiopulmonary bypass circuit and can be used for a number of applicatinos including arterial blood pumping, cardioplegia delivery, suction, and venting.",
      "report_date": "20200219",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "recall_number": "Z-1110-2020",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "84663",
      "termination_date": "20200512",
      "more_code_info": "",
      "recall_initiation_date": "20200110",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}