  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Plymouth",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98157",
      "recalling_firm": "Medtronic Inc",
      "address_1": "4600 Nathan Ln N",
      "address_2": "",
      "postal_code": "55442-2890",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan,  Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain,  Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam.",
      "recall_number": "Z-1098-2026",
      "product_description": "Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090  AB9G12100090  AB9G14080090  AB9G12080090  AB9U12060090  AB9U14100090  AB9U14080090  AB9U12080090  AB9U14060090  AB9G14100090  AB9G12060090",
      "product_quantity": "442",
      "reason_for_recall": "Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.",
      "recall_initiation_date": "20251204",
      "center_classification_date": "20260115",
      "report_date": "20260121",
      "code_info": "UDI-DI (GTIN):  00763000547332 00763000547271 00763000547318 00763000547264 00643169796225 00643169796300 00643169796294 00643169796232 00643169796287 00763000547325 00643169796300 00763000547257  Serial/Lot numbers  C112511 C113253 C113795 C117416 C117482 C121084 C123953 C133327 C133329 C134478 C138060 C150325",
      "more_code_info": ""
    }
  ]
}