      "status": "Ongoing",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "96083",
      "recalling_firm": "Getinge Usa Sales Inc",
      "address_1": "1 Geoffrey Way",
      "address_2": "",
      "postal_code": "07470-2035",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic distribution to Kentucky and Connecticut.",
      "recall_number": "Z-1074-2025",
      "product_description": "Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285.  Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.",
      "product_quantity": "6 units",
      "reason_for_recall": "Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.",
      "recall_initiation_date": "20250115",
      "center_classification_date": "20250207",
      "report_date": "20250219",
      "code_info": "Model Number: 6682285. UDI-DI: 07325710001141. Serial Numbers: 3761, 7635, 7633, 7630, 7632, 7631.",
      "more_code_info": ""
    }
  ]
}