{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Bryan",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89990",
      "recalling_firm": "Daavlin Distributing Company",
      "address_1": "205 W Bement St",
      "address_2": "N/A",
      "postal_code": "43506-1264",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SD, TN, TX, VA, WA, WI, WV and the countries of  Australia, Canada.",
      "recall_number": "Z-1074-2022",
      "product_description": "4 Series CX phototherapy units equipped with Daavlin's ClearLink Control System software versions v3.03h and 3.04v.  M4 Ser CX 311-20 120-240V 50/60Hz, item # 807QS0020CX6.",
      "product_quantity": "1",
      "reason_for_recall": "Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0\" inches to \"9. Resulting in over or under dosing.",
      "recall_initiation_date": "20220426",
      "center_classification_date": "20220512",
      "report_date": "20220518",
      "code_info": "UDI-DI: +B2809807QS0020CX60L. Serial Number: QS11571PLNBCX0020"
    }
  ]
}