{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Best",
      "state": "",
      "country": "Netherlands",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98109",
      "recalling_firm": "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.",
      "address_1": "Veenpluis 4-6",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.",
      "recall_number": "Z-1066-2026",
      "product_description": "Allura Xper FD20; Model Number: 722006;",
      "product_quantity": "143 units",
      "reason_for_recall": "Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.",
      "recall_initiation_date": "20251215",
      "center_classification_date": "20260113",
      "report_date": "20260121",
      "code_info": "Model Number: 722006; System Product Name: Allura Xper FD20; UDI-DI: N/A; Serial Numbers: 1370, 1376, 225, 249, 1273, 1496, 707, 1466, 1141, 1201, 359, 105, 285, 1433, 822, 1747, 318, 797, 124, 1518, 384, 1045, 328, 302, 1536, 514, 910, 2123, 1360, 1381, 149, 379, 387, 1733, 197, 314, 888, 189, 310, 209, 863, 325, 378, 2110, 166, 118, 1978, 304, 1696, 2524, 248, 1928, 294, 804, 1227, 56, 245, 1560, 1795, 902, 1295, 1123, 1424, 1958, 40, 1222, 1947, 2289, 1687, 2304, 666, 2086, 958, 1801, 2546, 1317, 1478, 589, 2154, 1557, 1580, 519, 273, 157, 1504, 836, 1365, 393, 2428, 1902, 2433, 1521, 115, 2023, 1003, 1509, 1494, 2171, 2391, 867, 1513, 29, 951, 1814, 284, 482, 455, 1763, 1899, 1968, 837, 856, 1537, 93, 1561, 1746, 1232, 430, CV.797, 1974, 507, 1949, 1275, 1427, 1967, 374, 116, CV.507, 7567201, 2225, 267, 931, 344, 745, 233, 506, 2230, 1864, 457, 280, 1556, 640, 2;",
      "more_code_info": ""
    }
  ]
}