{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mettawa",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93821",
      "recalling_firm": "Vyaire Medical",
      "address_1": "26125 N Riverwoods Blvd",
      "address_2": "",
      "postal_code": "60045-3420",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa",
      "recall_number": "Z-1064-2024",
      "product_description": "AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017",
      "product_quantity": "1,020,330 units",
      "reason_for_recall": "Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.",
      "recall_initiation_date": "20240110",
      "center_classification_date": "20240224",
      "report_date": "20240306",
      "code_info": "UDI/DI Case: 50190752114294, Each: 10190752114296; All manufacturing dates of 2017 and prior, and any products without a manufacturing date.",
      "more_code_info": ""
    }
  ]
}