      "status": "Ongoing",
      "city": "El Segundo",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89880",
      "recalling_firm": "Karl Storz Endoscopy",
      "address_1": "2151 E Grand Ave",
      "address_2": "N/A",
      "postal_code": "90245-5017",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA,WI, WV, and WY.  O.U.S.: Mexico, Lebanon, and South Korea",
      "recall_number": "Z-1061-2022",
      "product_description": "CYSTO-URETHRO-FIBERSCOPE REF 11272C1    CYSTOSCOPE, 15FR X 370MM, ADULT  REF 11272C2 and Kit Number 11272CK2    CYSTO-URETHRO-FIBERSCOPE  REF 11272CU1 and Kit Number 11272CUK1",
      "product_quantity": "All Serial Numbers manufactured/distributed since January 2018",
      "reason_for_recall": "Failure to achieve the expected six-log reduction in microorganisms following the disinfection process.",
      "recall_initiation_date": "20220401",
      "center_classification_date": "20220512",
      "report_date": "20220518",
      "code_info": "11272C1 n/a Flexible Cysto-Urethroscope Fiberscope Z18449US-BD (08-2018) UDI Code: 4048551226100; 11272C2 11272CK2 Flexible Cystoscope Z18449US-BD (08-2018) UDI Code: 4048551226100; 11272CU1 11272CUK1 Flexible Cystoscope Z18449US-BD (08-2018) UDI Code: 4048551226162;  All Serial Numbers manufactured/distributed since January 2018."
    }
  ]
}